Risk Factors Associated with Measurement Error across Different Types of Commonly Used Thermometers: A Cross-Sectional study from Ethiopia

Authors

  • Filmon Tsegay
  • Amsalu Bekele
  • Eyob Kebede Etissa
  • Eyob Woldeyohannes
  • Muralidhar M Kulkarni
  • Dominick Shaw
  • Colin Crooks
  • Andrew W Fogarty

DOI:

https://doi.org/10.20372/ejhd.v40i2.6877

Abstract

Abstract Background: No studies have compared infrared skin and sublingual thermometers to explore which factors may modify temperature measurement error. Objectives: To identify the measurement error with thermometers used in Ethiopia and the factors that may modify these errors. Methods: An observational study design was used. The study population comprised individuals attending a teaching hospital in Ethiopia. Data were collected on sublingual and infrared skin temperatures and compared with axillary temperature measurements. Results: Data were available from 2118 individuals (mean age of 43.2 years; 1310 (62%) were male). Among these individuals, 384 (18%) had temperature reading that differed by 0.5 °C or more from the axillary temperature when measured using the infrared thermometer. The corresponding figure for the sublingual thermometer was 365 (17%) individuals. Both thermometers showed an inverse association between mean temperature relative to the reference axillary temperature and measurement error (p<0.001 for both regression models). For the infrared skin thermometers, post-menopausal women who had a positive measurement error (+0.621; 95% CI: confidence intervals: +0.462 to +0.781), while a negative measurement error was associated with pregnancy, recent consumption of a hot or cold drink, higher body mass index, ambient relative humidity (-0.571°C; 95%CI: -0.624 to -0.518; highest tertile vs. lowest) and taking beta-blockers (-0.351°C; 95% CI: -0.416 to -0.286). Some of these modifying factors were also evident for sublingual thermometers, including beta-blockers use (-0.108°C; 95% CI: -0.163 to -0.052) and ambient humidity. Discussion: As sepsis accounts for approximately 20% of all global deaths, these observations should inform further studies in this area, with a particular consideration given to factors such as beta-blockers and ambient humidity. These data also raise concerns about the use of infrared skin temperature measurements in post-menopausal women. Conclusion: The measurement errors of infra-red skin and sublingual thermometers are modified by several risk factors, many of which modify peripheral circulation. [Ethiop. J. Health Dev. 2026; 40(2)] Keywords: Body temperature, Measurement error, thermometers, sepsis, hypotension, shock

Downloads

Published

2026-07-14

How to Cite

Filmon Tsegay, Amsalu Bekele, Eyob Kebede Etissa, Eyob Woldeyohannes, Muralidhar M Kulkarni, Dominick Shaw, Colin Crooks, & Andrew W Fogarty. (2026). Risk Factors Associated with Measurement Error across Different Types of Commonly Used Thermometers: A Cross-Sectional study from Ethiopia. The Ethiopian Journal of Health Development, 40(2). https://doi.org/10.20372/ejhd.v40i2.6877

Issue

Section

Original Articles

Most read articles by the same author(s)